7.10.5 If the displayed value is less than “200” (decimal point not
shown), a more accurate reading may be obtained by setting
the RANGE switch to a lower range. Rotate the RANGE switch
to the next lower position and observe the reading on the LCD
display. If the RANGE switch is already on the lowest position
(i.e. 2n F), no greater measurement resolution can be obtained.
Notes:
a)Onthe2nFand20nFranges,themetermayindicateresidual
capacitance.i.eThemeterreadingwillnotgotozero.Thisis
causedbytheinternalcapacitanceofthemetercircuitryorthe
testleads.
b)TheTestLeadsmayintroduceerrorwhenusedonthelow
capacitanceranges(like2nFand20nF).Tominimizetheerror,
notethedifferenceinreadingwiththeleadsinsertedintothe
meterandwiththeleadsremovedfromthemeter,andwhen
usingthetestleads,subtractthenoteddifferencefromthe
displayedreading.