23
Frequency Range: 50 to 400Hz
Overload Protection: 600V CAT IV AC/DC
1000V CAT III DC
700V CAT III AC
DC Current
Range Resolution Accuracy
400.0uA 0.1uA
4000uA 1uA
± 1.0% of rdg ± 4 dgts
40.00mA 10uA
400.0mA 100uA
± 1.2% of rdg ± 4 dgts
4.000A 1mA
10.00A 10mA
± 2.0% of rdg ± 5 dgts
Overload Protection: 500mA /660V and 10A / 600V fuse
Maximum Input: 400mA AC/DC on uA / mA ranges, 10A
AC/DC on 10A range (30 seconds max every 15 minutes).
AC Current
Range Resolution Accuracy
400.0uA 0.1uA
4000uA 1uA
± 1.2% of rdg ± 4 dgts
40.00mA 10uA
400.0mA 100uA
± 1.5% of rdg ± 4 dgts
4.000A 1mA
10.00A 10mA
± 2.0% of rdg ± 5 dgts
Overload Protection: 500mA /660V and 10A / 600V Fuse