6.1.11 Power Requirements
Batteries: Two, 9 volt alkaline batteries (NEDA 1604)
One, 1.5 volt alkaline D-cell (NEDA 13F)
Expected Life: Normal operation: 320 hours.
Sleep mode: 2000 hours.
Cutoff Voltage: 10.5 volts minimum for both 9V batteries
in series.
The ohmmeter circuit is designed to
properly zero with a D-cell battery from
1.0 to 1.8 volts.
Sleep Timer: Test set falls asleep after 3 to 7 minutes
with no meter deflection greater than 1%
of full scale typical.
6.1.12 Meter Movement
Full scale: 100 uA ±1%
Balance: ±1.0% of full scale.
Construction: Taut band construction, shock mounted.
100 degree deflection, critically damped.
Mirrored. High strength magnet.